Crosswalk: ISO 14971:2007 Vs. ISO 14971:20XX
Executive Summary
Here's a table cross-referencing the current 2007 version of international risk management standard ISO 14971 and its revised draft version, which will likely be published by the International Organization for Standardization (ISO) later this year.
Below is a crosswalk for the clauses and subclauses in the current version of international risk management standard ISO 14971 and its revised draft version, which will likely be published by the International Organization for Standardization (ISO) in the third quarter of 2019. (Check out our story about the ISO 14971 redo here.)
ISO 14971:2007 |
ISO 14971:20XX |
1 Scope |
1 Scope |
(New clause) |
2 Normative references |
2 Terms and definitions |
3 Terms and definitions |
2.1 accompanying document |
3.1 accompanying document |
(New definition) |
3.2 benefit |
2.2 harm |
3.3 harm |
2.3 hazard |
3.4 hazard |
2.4 hazardous situation |
3.5 hazardous situation |
2.5 intended use/intended purpose |
3.6 intended use/intended purpose |
2.6 in vitro diagnostic medical device/IVD medical device |
3.7 in vitro diagnostic medical device/IVD medical device |
2.7 life-cycle |
3.8 life-cycle |
2.8 manufacturer |
3.9 manufacturer |
2.9 medical device |
3.10 medical device |
2.10 objective evidence |
3.11 objective evidence |
2.11 post-production |
3.12 post-production |
2.12 procedure |
3.13 procedure |
2.13 process |
3.14 process |
(New definition) |
3.15 reasonably foreseeable misuse |
2.14 record |
3.16 record |
2.15 residual risk |
3.17 residual risk |
2.16 risk |
3.18 risk |
2.17 risk analysis |
3.19 risk analysis |
2.18 risk assessment |
3.20 risk assessment |
2.19 risk control |
3.21 risk control |
2.20 risk estimation |
3.22 risk estimation |
2.21 risk evaluation |
3.23 risk evaluation |
2.22 risk management |
3.24 risk management |
2.23 risk management file |
3.25 risk management file |
2.24 safety |
3.26 safety |
2.25 severity |
3.27 severity |
(New definition) |
3.28 state of the art |
2.26 top management |
3.29 top management |
2.27 use error |
3.30 use error |
2.28 verification |
3.31 verification |
3 General requirements for risk management |
4 General requirements for risk management |
3.1 Risk management process |
4.1 Risk management process |
3.2 Management responsibilities |
4.2 Management responsibilities |
3.3 Qualification of personnel |
4.3 Qualification of personnel |
3.4 Risk management plan |
4.4 Risk management plan |
3.5 Risk management file |
4.5 Risk management file |
4 Risk analysis |
5 Risk analysis |
4.1 Risk analysis process |
5.1 Risk analysis process |
4.2 Intended use and identification of characteristics related to the safety of the medical device |
5.2 Intended use and reasonably foreseeable misuse 5.3 Identification of characteristics related to safety |
4.3 Identification of hazards |
5.4 Identification of hazards and hazardous situations |
4.4 Estimation of the risk(s) for each hazardous situation |
5.5 Risk estimation |
5 Risk evaluation |
6 Risk evaluation |
6 Risk control |
7 Risk control |
6.1 Risk reduction |
(Subclause deleted) |
6.2 Risk control option analysis |
7.1 Risk control option analysis |
6.3 Implementation of risk control measure(s) |
7.2 Implementation of risk control measures |
6.4 Residual risk evaluation |
7.3 Residual risk evaluation |
6.5 Risk/benefit analysis |
7.4 Benefit-risk analysis |
6.6 Risks arising from risk control measures |
7.5 Risks arising from risk control measures |
6.7 Completeness of risk control |
7.6 Completeness of risk control |
7 Evaluation of overall residual risk acceptability |
8 Evaluation of overall residual risk |
8 Risk management report |
9 Risk management review |
9 Production and post-production information |
10 Production and post-production activities 10.1 Information collection 10.2 Information review 10.3 Actions |
Annex A Rationale for requirements |
Annex A Rationale for requirements |
Annex B Overview of the risk management process for medical devices |
Annex B Overview of the risk management process for medical devices |
Annex C Questions that can be used to identify medical device characteristics that could impact on safety |
Moved to ISO/TR 24971 |
Annex D Risk concepts applied to medical devices |
Moved to ISO/TR 24971 |
Annex E Examples of hazards, foreseeable sequences of events and hazardous situations |
Annex C Fundamental risk concepts |
Annex F Risk management plan |
Moved to ISO/TR 24971 |
Annex G Information on risk management techniques |
Moved to ISO/TR 24971 |
Annex H Guidance on risk management for in vitro diagnostic medical devices |
Moved to ISO/TR 24971 |
Annex I Guidance on risk analysis process for biological hazards |
Moved to ISO/TR 24971 |
Annex J Information for safety and information about residual risk |
Moved to ISO/TR 24971 |
Source: ISO 14971:20XX