FIRMS MUST INCLUDE EMC TEST RESULTS AND CRITERIA WHEN DECLARING CONFORMANCE TO IEC 60601-1-2 STANDARD; FDA TO PUBLISH 60601 ADOPTION PROPOSAL IN COMING WEEKS

This article was originally published in The Gray Sheet

Executive Summary

Manufacturers declaring conformity to the International Electrotechnical Commission's 60601-1-2 electromagnetic compatibility standard must include data from the EMC testing performed, FDA says in an Oct. 10 draft Blue Book memo to Office of Device Evaluation review staff. The memo is intended to provide ODE reviewers guidance on the use of the IEC 60601 series of device safety standards during the evaluation of premarket submissions.
Advertisement
Advertisement
UsernamePublicRestriction

Register

MT008971

Ask The Analyst

Please Note: You can also Click below Link for Ask the Analyst
Ask The Analyst

Your question has been successfully sent to the email address below and we will get back as soon as possible. my@email.address.

All fields are required.

Please make sure all fields are completed.

Please make sure you have filled out all fields

Please make sure you have filled out all fields

Please enter a valid e-mail address

Please enter a valid Phone Number

Ask your question to our analysts

Cancel